User: Guest  Login
Less Searchfields
Simple search
Original title:
Development of a large-animal X-ray dark-field scanner
Translated title:
Entwicklung eines Röntgen-Dunkelfeld Scanners für Großtiere
Author:
Gromann, Lukas
Year:
2017
Document type:
Dissertation
Faculty/School:
Fakultät für Physik
Advisor:
Pfeiffer, Franz (Prof. Dr.)
Referee:
Pfeiffer, Franz (Prof. Dr.); Finley, Jonathan J. (Prof. Dr.)
Language:
en
Subject group:
PHY Physik
TUM classification:
PHY 820d; MED 280d
Abstract:
This PhD thesis documents the development and optimization of a novel, high-energy large-animal X-ray dark-field (XDF) imaging system which overcomes the limitations of currently established setups. As the main result the very first in-vivo XDF chest radiographs of a pig with a field-of-view of 32cm x 32cm were acquired in 40sec scan time with an effective imaging dose of 0.08mSv. This result is a milestone in the bench-to-bedside translation of XDF imaging, with first human studies now at reach...     »
Translated abstract:
Diese Promotionsarbeit beschreibt die Entwicklung eines neuartigen Hochenergie-Röntgen-Dunkelfeld-Scanners, welcher die Limitierungen bestehender Systeme überwindet und damit die Bildgebung von Großtieren ermöglicht. Als Hauptresultat werden die weltweit ersten Thorax-Dunkelfeld-Bilder von lebenden Schweinen präsentiert. Diese Bilder erreichten ein Sichtfeld von 32cm x 32cm und wurden in 40s Scanzeit mit einer effektiven Strahlendosis von gerade einmal 0.08mSv aufgenommen. Dieses Ergebnis ist ei...     »
WWW:
https://mediatum.ub.tum.de/?id=1377460
Date of submission:
13.09.2017
Oral examination:
11.12.2017
File size:
178056544 bytes
Pages:
173
Urn (citeable URL):
https://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:91-diss-20171211-1377460-1-8
Last change:
11.12.2018
 BibTeX