User: Guest  Login
Original title:
Development and test of a new scanning geometry for Computed Tomography
Translated title:
Entwicklung und Überprüfung einer neuen Scangeometrie für Computertomographie
Author:
de las Heras, Hugo
Year:
2009
Document type:
Dissertation
Faculty/School:
Fakultät für Physik
Advisor:
Paretzke, Herwig G. (Prof. Dr.)
Referee:
Schreckenbach, Klaus (Prof. Dr.); Paretzke, Herwig G. (Prof. Dr.)
Language:
en
Subject group:
PHY Physik
Keywords:
Medical Imaging, Computed Tomography, scanning geometry, prototype, x-ray dose
Translated keywords:
Medizinische Bildgebung, Computertomographie, Scangeometrie, Demonstrator, Röntgendosis
Abstract:
This work presents an alternative scanning geometry for Computed Tomography with Double Optimal Reading, called CT D’OR. With this geometry, the Radon data are acquired after the X-ray beam is collimated through a special mask, which enables an efficient data collection and may reduce the patient dose. We have demonstrated with a test-device that the scanning geometry CT D’OR acquires data for two independent and complementary reconstructions. Additionally, we have shown with a Monte Carlo simul...     »
Translated abstract:
Diese Arbeit präsentiert eine neue Scangeometrie für Computertomographie mit doppelter optimaler Ablesung, genannt CT D’OR (CT with Dual Optimum Reading). Mit dieser Geometrie sammelt man die Radon Daten, nachdem der Röntgenf¨acher durch eine Maske kollimiert worden ist. Diese Eigenschaft ermöglicht eine effiziente Datensammlung und verringert möglicherweise die Patientendosis. Durch eine Monte Carlo Simulation wird gezeigt, dass man mit CT D’OR Bilder gleichwertiger Qualität wie mit normaler CT...     »
WWW:
https://mediatum.ub.tum.de/?id=677903
Date of submission:
30.10.2008
Oral examination:
20.01.2009
Pages:
134
Urn (citeable URL):
https://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:91-diss-20081029-677903-1-2
Last change:
23.01.2009
 BibTeX