Abstract:
We propose in this study a novel PET detector concept as insert for simultaneous PET/MR imaging, using arrays of Silicon Photomultipliers (SiPMs) as photodetectors, read out by a data acquisition system based on sampling ADCs. A 2 x 2 LSO-SiPM detector array and four single channel LYSO-SiPM detectors have been evaluated and compared to a LSO-APD detector. A 17.9% energy resolution and a 1.4 ns time resolution have been measured. No degradation of these values could be detected when simultaneous...     »