New Brightness Enhanced Positron Microbeam for Spatially Resolved Doppler Broadening Spectroscopy at NEPOMUC
Translated title:
Ortsaufgelöste Doppler-Verbreiterungsspektroskopie an NEPOMUC mit einem neuen emittanzerhöhten Positronenmikrostrahl
Author:
Gigl, Thomas
Year:
2018
Document type:
Dissertation
Faculty/School:
Fakultät für Physik
Advisor:
Hugenschmidt, Christoph (Priv.-Doz. Dr.)
Referee:
Hugenschmidt, Christoph (Priv.-Doz. Dr.); Dollinger, Günther (Prof. Dr.)
Language:
en
Subject group:
PHY Physik
TUM classification:
PHY 600d; WER 000d; PHY 480d
Abstract:
Within the scope of this thesis, a new coincident Doppler broadening spectrometer (CDBS upgrade) was designed, established and successfully set into routine operation at the high-intensity positron source NEPOMUC. The implementation of a new positron microbeam enables non-destructive evaluation of the defect distribution and the chemical composition at the positron-electron annihilation site. Its capabilities have first been demonstrated on laser beam and friction stir welded AlCu alloys.
Translated abstract:
Im Rahmen dieser Arbeit wurde ein neues koinzidentes Dopplerverbreiterungs-Spektrometer (CDBS) an der hochintensiven Positronenquelle NEPOMUC entwickelt, aufgebaut und erfolgreich in Routinebetrieb genommen. Der CDBS-Mikrostrahl ermöglicht die lateral hochaufgelöste und tiefenabhängige Defektspektroskopie sowie die zerstörungsfreie Untersuchung der Fehlstellen- und Elementverteilungen am Annihilationsort. Dies konnte an Laser- und Rührreibschweißnähten einer AlCu Legierung demonstriert werden.