User: Guest  Login
Title:

Analysis of carcinogenic polycyclic aromatic hydrocarbons in complex environmental mixtures by LC-APPI-MS/MS.

Document type:
Comparative Study; Evaluation Studies; Journal Article; Research Support, Non-U.S. Gov't
Author(s):
Hutzler, C; Luch, A; Filser, JG
Abstract:
Here we developed a highly sensitive, fast and reliable liquid chromatography tandem mass spectrometry (LC-MS/MS) method for the detection and analysis of 16 different polycyclic aromatic hydrocarbons (PAHs) and nitro-PAHs that have been identified as carcinogens and classified according to their biological potency. Comparison to standard analysis procedures based on gas chromatography-mass spectrometry (GC-MS) instrumentation demonstrated an improved easiness of sample preparation and sensitivity of detection achieved with the new LC-MS/MS method employing an atmospheric pressure photoionization (APPI) source attached to an API 4000 mass spectrometer (LC-APPI-MS/MS). The favorable outcome could be confirmed by analyzing complex mixtures such as certain Standard Reference Materials (SRMs) obtained from the National Institute of Standards & Technology (NIST), i.e., SRM 1975 and SRM 2975, and several diesel exhaust soots provided by the German automobile industry. Certified concentrations of individual analytes provided by NIST not only could be confirmed, but additional extremely potent carcinogens such as several isomeric hexacyclic dibenzopyrenes (DBPs), 5-methylchrysene (5-MC), and others have been detected in these crude samples in a concentration range down to below 1 ng g(-1) raw material.
Journal title abbreviation:
Anal Chim Acta
Year:
2011
Journal volume:
702
Journal issue:
2
Pages contribution:
218-24
Language:
eng
Fulltext / DOI:
doi:10.1016/j.aca.2011.07.003
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/21839201
Print-ISSN:
0003-2670
TUM Institution:
Institut für Toxikologie und Umwelthygiene
 BibTeX