Benutzer: Gast  Login
Mehr Felder
Einfache Suche
Titel:

3D surface measurement for medical application--technical comparison of two established industrial surface scanning systems.

Dokumenttyp:
Comparative Study; Journal Article; Research Support, Non-U.S. Gov't
Autor(en):
Schwenzer-Zimmerer, K; Haberstok, J; Kovacs, L; Boerner, BI; Schwenzer, N; Juergens, P; Zeihofer, HF; Holberg, C
Abstract:
In 3D mapping of flexible surfaces (e.g. human faces) measurement errors due to movement or positioning occur. Aggravated by equipment- or researcher-caused mistakes considerable deviations can result. Therefore first the appliances' precision handling and reliability in clinical environment must be established. Aim of this study was to investigate accuracy and precision of two contact-free 3D measurement systems (white light vs. laser). Standard specimens of known diameter for sphere deviation, touch deviation and plane deviation were tested. Both systems are appropriate for medical application acquiring solid data (
Zeitschriftentitel:
J Med Syst
Jahr:
2008
Band / Volume:
32
Heft / Issue:
1
Seitenangaben Beitrag:
59-64
Sprache:
eng
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/18333407
Print-ISSN:
0148-5598
TUM Einrichtung:
Lehrstuhl für Plastische Chirurgie und Handchirurgie (Prof. Machens)
 BibTeX