Objectives: Point-of-care antigen tests (PoC-AgTs) for the rapid detection of SARS-CoV-2 infection enable screening of additional populations with less effort, independent of laboratories and at a low cost. PoC-AgTs have therefore been included in national testing strategies with additional quality requirements to address limitations in specificity and sensitivity. Information given in the package inserts of the test providers should enable the user to evaluate the performance of a PoC-AgT in advance. The quality of this information has been independently assessed since the Corona Test Ordinance came into force in Germany in October 2020.
Methods: The completeness of analytical and diagnostic performance specifications was assessed for all package inserts publicly available via the Paul Ehrlich Institute (PEI). It was ascertained whether the minimum criteria, recommendations, and extended criteria of the PEI were comprehensibly fulfilled. The number of tests removed from the list by March 2021 was determined.
Results: By the closing date of the survey (17.11.2020), the PEI had listed 165 PoC-AgTs that formally fulfilled the minimum criteria and were thus reimbursed. A total of 78 identical systems were identified. Almost all providers were found to have gaps in the information on the validation results of their tests, meaning that an evaluation of performance is only possible to a limited extent. Until March 2021, 25 non-identical PoC-AgTs have been removed from the list.
Conclusions: Many PoC-AgTs could not be comprehensively evaluated based on the information provided by the provider. Users are therefore dependent on providerindependent sources of information.
«
Objectives: Point-of-care antigen tests (PoC-AgTs) for the rapid detection of SARS-CoV-2 infection enable screening of additional populations with less effort, independent of laboratories and at a low cost. PoC-AgTs have therefore been included in national testing strategies with additional quality requirements to address limitations in specificity and sensitivity. Information given in the package inserts of the test providers should enable the user to evaluate the performance of a PoC-AgT in ad...
»