Accuracy of a hand-held resistance-free flowmeters for flow adjustments of Hirst-Type pollen traps

Dokumenttyp:
Article
Autor(en):
Suarez-Suarez, Mariel; Maya-Manzano, Jose M. M.; Clot, Bernard; Graber, Marie-Jose; Sallin, Christine; Tummon, Fiona; Buters, Jeroen
Abstract:
The Hirst-type pollen trap (1952) is the most common device for aerobiological measurements of pollen and fungal spores in ambient air. In the 1960s the in-line flowmeter was removed and studies since then showed considerable variability in the airflow pumped through the instrument when using hand-held rotameters with an internal airflow resistance to adjust flowrates. To avoid this problem, our study compared the variability of airflow rates of Hirst-type traps when using commercially-available...     »