Background: This study compares two latest-generation self-expanding transcatheter heart valves (THV), the supra-annular ACURATE neo (Boston Scientific) and the intra-annular Centera THV (Edwards Lifesciences).
Methods: In this single center observational cohort trial 317 patients treated with the ACURATE neo and 78 patients treated with the Centera TVH were included. The main endpoints were device success and the early safety endpoint at 30 days.
Results: Besides higher incidence of diabetes mellitus and higher body mass index in patients treated with the ACURATE neo THV, there were no baseline differences between the groups. Device success was similar in both groups (neo: 91.8% vs. Centera: 93.6%; P=0.598), with numerically higher rates of moderate to severe paravalvular leakage in the ACURATE neo group (4.7% vs. 1.3%; P=0.214). At 30 days all-cause mortality rates were low in both groups (0.3% vs. 0%; P=0.620) and no difference occurred in the early safety at 30 days (19.3% vs. 16.7%; P=0.599). However, all-cause stroke rates were significantly higher in patients treated with the Centera THV (6.4 vs. 1.6%; P=0.015).
Conclusions: The ACURATE neo and the Centera THV show low mortality rates as well as comparable, favorable hemodynamics. The finding of higher stroke rates at 30 days with the repositionable Centera SE-THV needs further assessment.
«
Background: This study compares two latest-generation self-expanding transcatheter heart valves (THV), the supra-annular ACURATE neo (Boston Scientific) and the intra-annular Centera THV (Edwards Lifesciences).
Methods: In this single center observational cohort trial 317 patients treated with the ACURATE neo and 78 patients treated with the Centera TVH were included. The main endpoints were device success and the early safety endpoint at 30 days.
Results: Besides higher incidence of diabetes me...
»