Multicenter evaluation of the analytical and clinical performance of the Elecsys S100 immunoassay in patients with malignant melanoma.

Dokumenttyp:
Evaluation Studies; Journal Article; Multicenter Study; Article
Autor(en):
Alber, B; Hein, R; Garbe, C; Caroli, U; Luppa, PB
Abstract:
The aim of this multicenter study was to evaluate the technical and clinical performance of the Elecsys S100 electrochemiluminescence immunoassay and to assess its utility as a tumor marker in patients with malignant melanoma. Imprecision studies yielded within-run coefficients of variation (CVs) of 0.7-2.0% and between-day CVs of 1.0-6.4%. Serum samples that were distributed to 11 participating laboratories for a comparability analysis resulted in excellent recoveries of 93-105% related to the...     »