Benutzer: Gast  Login
Mehr Felder
Einfache Suche
Titel:

Ultrawideband reflection-mode optoacoustic mesoscopy.

Dokumenttyp:
Journal Article; Research Support, Non-U.S. Gov't; Article
Autor(en):
Omar, Murad; Soliman, Dominik; Gateau, Jérôme; Ntziachristos, Vasilis
Abstract:
We developed a reflection-mode optoacoustic mesoscopy system, based on raster-scanning of a custom designed spherically focused ultrasound detector, enabling seamless epi-illumination of the volume imaged. We study the performance of acoustic-resolution mesoscopy operating at an ultrawideband bandwidth of 20-180 MHz. i.e., a frequency band spreading over virtually an order of magnitude. Using tomographic reconstruction we showcase previously unreported, to our knowledge, axial resolutions of 4 ?...     »
Zeitschriftentitel:
Opt Lett
Jahr:
2014
Band / Volume:
39
Heft / Issue:
13
Seitenangaben Beitrag:
3911-4
Sprache:
eng
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/24978769
Print-ISSN:
0146-9592
TUM Einrichtung:
Lehrstuhl für Biologische Bildgebung - Zusammenarbeit mit dem Helmholtz-Zentrum München (Prof. Ntziachristos)
 BibTeX