User: Guest  Login
Title:

Measurements to predict the time of target replacement of a helical tomotherapy.

Document type:
Journal Article; Research Support, Non-U.S. Gov't
Author(s):
Kampfer, S; Schell, S; Duma, MN; Wilkens, JJ; Kneschaurek, P
Abstract:
Intensity-modulated radiation therapy (IMRT) requires more beam-on time than normal open field treatment. Consequently, the machines wear out and need more spare parts. A helical tomotherapy treatment unit needs a periodical tungsten target replacement, which is a time consuming event. To be able to predict the next replacement would be quite valuable. We observed unexpected variations towards the end of the target lifetime in the performed pretreatment measurements for patient plan verification...     »
Journal title abbreviation:
J Appl Clin Med Phys
Year:
2011
Journal volume:
12
Journal issue:
4
Pages contribution:
3596
Language:
eng
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/22089018
TUM Institution:
Klinik und Poliklinik für RadioOnkologie und Strahlentherapie
 BibTeX