Benutzer: Gast  Login
Titel:

Measurements to predict the time of target replacement of a helical tomotherapy.

Dokumenttyp:
Journal Article; Research Support, Non-U.S. Gov't
Autor(en):
Kampfer, S; Schell, S; Duma, MN; Wilkens, JJ; Kneschaurek, P
Abstract:
Intensity-modulated radiation therapy (IMRT) requires more beam-on time than normal open field treatment. Consequently, the machines wear out and need more spare parts. A helical tomotherapy treatment unit needs a periodical tungsten target replacement, which is a time consuming event. To be able to predict the next replacement would be quite valuable. We observed unexpected variations towards the end of the target lifetime in the performed pretreatment measurements for patient plan verification...     »
Zeitschriftentitel:
J Appl Clin Med Phys
Jahr:
2011
Band / Volume:
12
Heft / Issue:
4
Seitenangaben Beitrag:
3596
Sprache:
eng
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/22089018
TUM Einrichtung:
Klinik und Poliklinik für RadioOnkologie und Strahlentherapie
 BibTeX