Benutzer: Gast  Login
Weniger Felder
Einfache Suche
Titel:

Evaluation of the attenuation properties of MR equipment for its use in a whole-body PET/MR scanner.

Dokumenttyp:
Evaluation Studies; Journal Article
Autor(en):
Delso, G; Martinez-Möller, A; Bundschuh, RA; Ladebeck, R; Candidus, Y; Faul, D; Ziegler, SI
Abstract:
The combination of magnetic resonance imaging (MR) and positron emission tomography (PET) scanners can provide a powerful tool for clinical diagnosis and investigation. Among the challenges of developing a combined scanner, obtaining attenuation maps for PET reconstruction is of critical importance. This requires accounting for the presence of MR hardware in the field of view. The attenuation introduced by this hardware cannot be obtained from MR data. We propose the creation of attenuation mode...     »
Zeitschriftentitel:
Phys Med Biol
Jahr:
2010
Band / Volume:
55
Heft / Issue:
15
Seitenangaben Beitrag:
4361-74
Sprache:
eng
Volltext / DOI:
doi:10.1088/0031-9155/55/15/011
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/20647598
Print-ISSN:
0031-9155
TUM Einrichtung:
Klinik und Poliklinik für Nuklearmedizin
 BibTeX