PURPOSE: To develop a Dixon-based B 0 $$ {\mathrm{B}}_0 $$ self-navigation approach to estimate and correct temporal B 0 $$ {\mathrm{B}}_0 $$ variations in radial stack-of-stars gradient echo imaging for quantitative body MRI.
METHODS: The proposed method estimates temporal B 0 $$ {\mathrm{B}}_0 $$ variations using a B 0 $$ {\mathrm{B}}_0 $$ self-navigator estimated by a graph-cut-based water-fat separation algorithm on the oversampled k-space center. The B 0 $$ {\mathrm{B}}_0 $$ self-navigator was employed to correct for phase differences between radial spokes (one-dimensional [1D] correction) and to perform a motion-resolved reconstruction to correct spatiotemporal pseudo-periodic B 0 $$ {\mathrm{B}}_0 $$ variations (three-dimensional [3D] correction). Numerical simulations, phantom experiments and in vivo neck scans were performed to evaluate the effects of temporal B 0 $$ {\mathrm{B}}_0 $$ variations on the field-map, proton density fat fraction (PDFF) and T 2 ∗ $$ {\mathrm{T}}_2^{\ast } $$ map, and to validate the proposed method.
RESULTS: Temporal B 0 $$ {\mathrm{B}}_0 $$ variations were found to cause signal loss and phase shifts on the multi-echo images that lead to an underestimation of T 2 ∗ $$ {\mathrm{T}}_2^{\ast } $$ , while PDFF mapping was less affected. The B 0 $$ {\mathrm{B}}_0 $$ self-navigator captured slowly varying temporal B 0 $$ {\mathrm{B}}_0 $$ drifts and temporal variations caused by respiratory motion. While the 1D correction effectively corrected B 0 $$ {\mathrm{B}}_0 $$ drifts in phantom studies, it was insufficient in vivo due to 3D spatially varying temporal B 0 $$ {\mathrm{B}}_0 $$ variations with amplitudes of up to 25 Hz at 3 T near the lungs. The proposed 3D correction locally improved the correction of field-map and T 2 ∗ $$ {\mathrm{T}}_2^{\ast } $$ and reduced image artifacts.
CONCLUSION: Temporal B 0 $$ {\mathrm{B}}_0 $$ variations particularly affect T 2 ∗ $$ {\mathrm{T}}_2^{\ast } $$ mapping in radial stack-of-stars imaging. The self-navigation approach can be applied without modifying the MR acquisition to correct for B 0 $$ {\mathrm{B}}_0 $$ drift and physiological motion-induced B 0 $$ {\mathrm{B}}_0 $$ variations, especially in the presence of fat.
«