Transpulmonary thermodilution (TPTD)-derived global end-diastolic volume index (GEDVI) is a static marker of preload which better predicted volume responsiveness compared to filling pressures in several studies. GEDVI can be generated with at least two devices: PiCCO and EV-1000. Several studies showed that uncorrected indicator injection into a femoral central venous catheter (CVC) results in a significant overestimation of GEDVI by the PiCCO-device. Therefore, the most recent PiCCO-algorithm corrects for femoral indicator injection. However, there are no systematic data on the impact of femoral indicator injection for the EV-1000 device. Furthermore, the correction algorithm of the PiCCO is poorly validated. Therefore, we prospectively analyzed 14 datasets from 10 patients with TPTD-monitoring undergoing central venous catheter (CVC)- and arterial line exchange. PiCCO was replaced by EV-1000, femoral CVCs were replaced by jugular/subclavian CVCs and vice-versa. For PiCCO, jugular and femoral indicator injection derived GEDVI was comparable when the correct information about femoral catheter site was given (p = 0.251). By contrast, GEDVI derived from femoral indicator injection using the EV-1000 was obviously not corrected and was substantially higher than jugular GEDVI measured by the EV-1000 (846 ± 250 vs. 712 ± 227 ml/m2; p = 0.001). Furthermore, measurements of GEDVI were not comparable between PiCCO and EV-1000 even in case of jugular indicator injection (p = 0.003). This is most probably due to different indexations of the raw value GEDV. EV-1000 could not be recommended to measure GEDVI in case of a femoral CVC. Furthermore, different indexations used by EV-1000 and PiCCO should be considered even in case of a jugular CVC when comparing GEDVI derived from PiCCO and EV-1000.
«
Transpulmonary thermodilution (TPTD)-derived global end-diastolic volume index (GEDVI) is a static marker of preload which better predicted volume responsiveness compared to filling pressures in several studies. GEDVI can be generated with at least two devices: PiCCO and EV-1000. Several studies showed that uncorrected indicator injection into a femoral central venous catheter (CVC) results in a significant overestimation of GEDVI by the PiCCO-device. Therefore, the most recent PiCCO-algorithm c...
»