Sustained low efficiency dialysis should not be interrupted for performing transpulmonary thermodilution measurements.

Dokumenttyp:
Journal Article
Autor(en):
Geith, Stefanie; Stecher, Lynne; Rabe, Christian; Sack, Stefan; Eyer, Florian
Abstract:
Treatment of multiple organ failure frequently requires enhanced hemodynamic monitoring. When renal replacement is indicated, it remains unclear whether transpulmonary thermodilution (TPTD) measurements are influenced by renal replacement therapy (RRT) and whether RRT should be paused for TPTD measurements. Our aim was therefore to investigate the effect of pausing RRT on TPTD results in two dialysis catheter locations.In total, 62 TPTD measurements in 24 patients (APACHE: 32 ± 7 [mean ± standar...     »