User: Guest  Login
Title:

Energy-Dependent Neutron Imaging with a Double Crystal Monochromator at the {{ANTARES}} Facility at {{FRM II}}

Document type:
Zeitschriftenaufsatz
Author(s):
Schulz, M.; B̈oni, P.; Calzada, E.; M̈uhlbauer, M.; Schillinger, B.
Abstract:
A double crystal-type monochromator has been installed at the ANTARES facility at FRM II, Munich, close to the collimator where the beam size is still only a few centimetres. Using this new set-up, it is now possible to perform high-resolution monochromatic neutron imaging in a wavelength range from 2.7 to 6.5 A˚ with a wavelength resolution Dl/l of 4\textendash{}1%. Most of the Bragg edges of materials of interest for neutron radiography lie in this wavelength region. Exposure times in the orde...     »
Journal title:
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Year:
2009
Journal volume:
605
Month:
jun
Journal issue:
1-2
Pages contribution:
33
Language:
en
Fulltext / DOI:
doi:10.1016/j.nima.2009.01.123
Print-ISSN:
01689002
 BibTeX