User: Guest  Login
Document type:
Journal Article
Author(s):
Huisman, MC; Reder, S; Weber, AW; Ziegler, SI; Schwaiger, M
Title:
Performance evaluation of the Philips MOSAIC small animal PET scanner.
Abstract:
PURPOSE: In this study an evaluation of the performance of the Philips MOSAIC small animal PET scanner is presented, with special emphasis on the ability of the system to provide quantitatively accurate PET images. METHODS: The performance evaluation was structured according to NEMA-like procedures. RESULTS: The transaxial spatial resolution of the system (radial component) ranged between 2.7 mm FWHM at the centre and 3.2 mm FWHM at a radial offset of 45 mm from the centre. The axial spatial res...     »
Journal title abbreviation:
Eur J Nucl Med Mol Imaging
Year:
2007
Journal volume:
34
Journal issue:
4
Pages contribution:
532-40
Language:
eng
Fulltext / DOI:
doi:10.1007/s00259-006-0271-7
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/17119959
Print-ISSN:
1619-7070
TUM Institution:
Klinik und Poliklinik für Nuklearmedizin
 BibTeX