Benutzer: Gast  Login
Titel:

Development and clinical evaluation of an ionization chamber array with 3.5 mm pixel pitch for quality assurance in advanced radiotherapy techniques.

Dokumenttyp:
Evaluation Studies; Journal Article; Article
Autor(en):
Togno, M; Wilkens, J J; Menichelli, D; Oechsner, M; Perez-Andujar, A; Morin, O
Abstract:
To characterize a new air vented ionization chamber technology, suitable to build detector arrays with small pixel pitch and independence of sensitivity on dose per pulse.The prototype under test is a linear array of air vented ionization chambers, consisting of 80 pixels with 3.5 mm pixel pitch distance and a sensitive volume of about 4 mm(3). The detector has been characterized with (60)Co radiation and MV x rays from different linear accelerators (with flattened and unflattened beam qualities...     »
Zeitschriftentitel:
Med Phys
Jahr:
2016
Band / Volume:
43
Heft / Issue:
5
Seitenangaben Beitrag:
2283
Sprache:
eng
Volltext / DOI:
doi:10.1118/1.4945414
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/27147340
Print-ISSN:
0094-2405
TUM Einrichtung:
Klinik und Poliklinik für RadioOnkologie und Strahlentherapie
 BibTeX