Rapid single-scan T2*-mapping using exponential excitation pulses and image-based correction for linear background gradients.

Document type:
Evaluation Studies; Journal Article; Research Support, Non-U.S. Gov't
Author(s):
Baudrexel, S; Volz, S; Preibisch, C; Klein, JC; Steinmetz, H; Hilker, R; Deichmann, R
Abstract:
A method for fast quantitative T(2)* mapping based on multiple gradient-echo (multi-GE) imaging with correction for static magnetic field inhomogeneities is described, using an exponential excitation pulse. Field gradient maps are obtained from the phase information and modulus data are subsequently corrected, allowing for simple monoexponential T(2)* fitting. Echoes with long echo times suffering from major signal losses due to field inhomogeneities are excluded from the analysis. The acquisiti...     »
Journal title abbreviation:
Magn Reson Med
Year:
2009
Journal volume:
62
Journal issue:
1
Pages contribution:
263-8
Language:
eng
Fulltext / DOI:
doi:10.1002/mrm.21971
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/19353655
Print-ISSN:
0740-3194
TUM Institution:
Fachgebiet Neuroradiologie (Prof. Zimmer)