User: Guest  Login
Title:

Comparison of different excimer laser ablation frequencies (50, 200, and 500 Hz).

Document type:
Comparative Study; Journal Article
Author(s):
Winkler von Mohrenfels, C; Khoramnia, R; Lohmann, CP
Abstract:
PURPOSE: The aim of this study was to compare the influence of different excimer laser (193 nm) ablation frequencies (50, 200, and 500 Hz) on porcine cornea. METHODS: For this study, fresh porcine cadaver eyes were obtained from the butcher. PRK and LASIK were performed using a WaveLight Concept System 500 excimer laser. Various repetition rates (50, 200, and 500 Hz) and intended corrections (between -3 and -9 D) were used. Corneal endothelium was investigated pre- and post-operatively with the...     »
Journal title abbreviation:
Graefes Arch Clin Exp Ophthalmol
Year:
2009
Journal volume:
247
Journal issue:
11
Pages contribution:
1539-45
Language:
eng
Fulltext / DOI:
doi:10.1007/s00417-009-1102-x
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/19597838
Print-ISSN:
0721-832X
TUM Institution:
Klinik und Poliklinik für Augenheilkunde
 BibTeX