Abstract:
PURPOSE: In this study an evaluation of the performance of the Philips MOSAIC small animal PET scanner is presented, with special emphasis on the ability of the system to provide quantitatively accurate PET images. METHODS: The performance evaluation was structured according to NEMA-like procedures. RESULTS: The transaxial spatial resolution of the system (radial component) ranged between 2.7 mm FWHM at the centre and 3.2 mm FWHM at a radial offset of 45 mm from the centre. The axial spatial res...     »