User: Guest  Login
Title:

Gating failure can result in underestimation of cardiac function in myocardial perfusion scintigraphy

Document type:
Zeitschriftenaufsatz
Author(s):
Villagran Asiares, Alberto; Yakushev, Igor; Nekolla, Stephan G.
Abstract:
Here, we present a case with a pacemaker due to an atrioventricular (AV) block 2 Mobitz type, in whom a gating failure resulted in a relevant underestimation of cardiac function in myocardial perfusion scintigraphy. A set of quality control steps for gating errors is proposed.
FP7 Project ID:
764458
Horizon 2020:
This work is part of a project that has received funding from the European Union’s Horizon 2020 research and innovation program under the Marie Skłodowska-Curie Grant Agreement No 764458.
Journal title:
Journal of Nuclear Cardiology
Year:
2020
Language:
en
Fulltext / DOI:
doi:10.1007/s12350-020-02430-8
Publisher:
Springer Science and Business Media LLC
E-ISSN:
1071-35811532-6551
Date of publication:
11.11.2020
Copyright statement:
 BibTeX