Ruge, Hendrik; Alvarez-Covarrubias, Hector A; Deutsch, Oliver; Alalawi, Zahra; Vitanova, Keti; Lange, Rüdiger
Title:
Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900.
Abstract:
Background: Bioprosthetic valve fracturing (BVF) results in low gradients following valve-in-valve transcatheter aortic valve replacement (ViV-TAVR). For the commonly used Edwards PERIMOUNT valve data from bench-testing are lacking to provide technical specifications for successful BVF during ViV-TAVR. Methods: Using four Perimount 19- and 21-mm valves, in-vitro high-pressure balloon valvuloplasty with the True Dilatation Balloon Valvuloplasty Catheter and Atlas Gold PTA Dilatation Catheter was...     »
Journal title abbreviation:
Front Cardiovasc Med
Year:
2022
Journal volume:
9
Fulltext / DOI:
doi:10.3389/fcvm.2022.859088
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/35774373
TUM Institution:
Klinik für Herz- und Gefäßchirurgie (Prof. Lange)
 BibTeX