PURPOSE: In current clinical practice, intraoperative repositioning of mobile C-arms is challenging due to a lack of visual cues and efficient guiding tools. This can be detrimental to the surgical workflow and lead to additional radiation burdens for both patient and personnel. To overcome this problem, we present our novel approach Lidar-based X-ray Positioning for Mobile C-arms (RAY-POS) for assisting circulating nurses during intraoperative C-arm repositioning without requiring external aids.
METHODS: RAY-POS consists of a localization module and a graphical user interface for guiding the user back to a previously recorded C-Arm position. We conducted a systematic comparison of simultaneous localization and mapping (SLAM) algorithms using different attachment positions of light detection and ranging (LIDAR) sensors to benchmark localization performance within the operating room (OR). For two promising combinations, we conducted further end-to-end repositioning tests within a realistic OR setup.
RESULTS: SLAM algorithm gmapping with a LIDAR sensor mounted 40 cm above the C-arm's horizontal unit performed best regarding localization accuracy and long-term stability. The distribution of the repositioning error yielded an effective standard deviation of 7.61 mm.
CONCLUSION: We conclude that a proof-of-concept for LIDAR-based C-arm repositioning without external aids has been achieved. In future work, we mainly aim at extending the capabilities of our system and evaluating the usability together with clinicians.
«
PURPOSE: In current clinical practice, intraoperative repositioning of mobile C-arms is challenging due to a lack of visual cues and efficient guiding tools. This can be detrimental to the surgical workflow and lead to additional radiation burdens for both patient and personnel. To overcome this problem, we present our novel approach Lidar-based X-ray Positioning for Mobile C-arms (RAY-POS) for assisting circulating nurses during intraoperative C-arm repositioning without requiring external aids...
»