A novel transcatheter aortic valve with a form-fitting anchor for self-alignment: feasibility in a chronic preclinical model.

Document type:
Journal Article
Author(s):
Noterdaeme, Timothée; Marx, Nikolaus; Lange, Rüdiger
Abstract:
OBJECTIVES: Our goal was to assess the viability of an alternative anchoring design for a transcatheter aortic valve based on a form-fitting principle with a self-expanding nitinol frame to reduce issues related to excess radial force. METHODS: A 26-mm outer diameter prototype of a self-expanding nitinol frame was developed to reduce the amount of necessary radial force by utilizing additional anchoring via protruding arms in each aortic sinus, thus allowing for a form-fitting principle as well...     »
Journal title abbreviation:
Interact Cardiovasc Thorac Surg
Year:
2019
Journal volume:
29
Journal issue:
1
Pages contribution:
8-14
Fulltext / DOI:
doi:10.1093/icvts/ivz005
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/30768173
Print-ISSN:
1569-9293
TUM Institution:
Klinik für Herz- und Gefäßchirurgie (Prof. Lange)