PURPOSE: The primary objective of the retrospective study was to collect speech intelligibility data on children and adolescents implanted with the vibrating ossicular prosthesis (VORP) 503.
METHODS: This was a retrospective, multicentre study on 55 children and adolescents from 6 German clinics aged between 5 and 17 years suffering from mixed or conductive hearing loss implanted with a VORP 503. Pre- and postoperative bone-conduction pure tone thresholds were measured at 0.5, 1, 2 and 4 kHz, and word recognition scores in the unaided and VORP 503-aided conditions using monosyllabic speech intelligibility tests measured at 65-dB sound pressure level (SPL) were determined.
RESULTS: Mean pre- and postoperative bone-conduction thresholds remained unchanged, showing the preservation of inner ear hearing. Speech intelligibility assessed in quiet at 65-dB SPL improved on average from 24.5% (SD ± 25.4) unaided to 86.4% (SD ± 13.4) aided. The average improvement of 61.9% (SD ± 25.3) was clinically and statistically significant. A total of three complications were found in the medical records of 55 subjects. The responsible investigators judged these events as procedure related.
CONCLUSION: The treatment of children suffering from conductive or mixed hearing loss with the VORP 503 implant demonstrates excellent aided benefit in terms of speech understanding and only minor complications.
«
PURPOSE: The primary objective of the retrospective study was to collect speech intelligibility data on children and adolescents implanted with the vibrating ossicular prosthesis (VORP) 503.
METHODS: This was a retrospective, multicentre study on 55 children and adolescents from 6 German clinics aged between 5 and 17 years suffering from mixed or conductive hearing loss implanted with a VORP 503. Pre- and postoperative bone-conduction pure tone thresholds were measured at 0.5, 1, 2 and 4 kHz, an...
»