Title:

Gating failure can result in underestimation of cardiac function in myocardial perfusion scintigraphy

Document type:
Zeitschriftenaufsatz
Author(s):
Villagran Asiares, Alberto; Yakushev, Igor; Nekolla, Stephan G.
Abstract:
Here, we present a case with a pacemaker due to an atrioventricular (AV) block 2 Mobitz type, in whom a gating failure resulted in a relevant underestimation of cardiac function in myocardial perfusion scintigraphy. A set of quality control steps for gating errors is proposed.