The Sapien valve provides enough grip to be implanted in pulmonary position without a pre-stent.

Dokumenttyp:
Article; Journal Article
Autor(en):
Tanase, Daniel; Georgiev, Stanimir; Eicken, Andreas; Ewert, Peter
Abstract:
Background: Pre-stenting is a widely used technique for transcatheter pulmonary valve implantation (TPVI). For the Melody valve a pre-stent creates a reliable and solid landing zone. In contrast to the Melody valve, the Edwards Sapien valve consists of a strong balloon expandable stent designed for implantation in the aortic position. The usage of a Sapien valve in pulmonary position might therefore obviate the need of a pre-stent. Methods: This retrospective, single centre study reviewed the pr...     »