Benutzer: Gast  Login
Titel:

[Experimental analysis of neutral, asymmetric and constraint liners for total hip replacement: investigation of range of motion and protection against joint instability]

Dokumenttyp:
Evaluation Studies; Journal Article; Validation Studies
Autor(en):
Bader, R; Steinhauser, E; Scholz, R; Simnacher, M; Mittelmeier, W
Abstract:
AIM: Recurrent dislocation after total hip replacement is a severe complication, which requires specific treatment and implants. The purpose of the present study was to compare a constraint liner with an elevated rim and standard liner regarding their range of motion and dislocation stability. METHOD: With a test device, range of motion until impingement (ROM (Imp)) and dislocation (ROM (Lux)) were experimentally analyzed using the above-mentioned insert types of a commercial total hip system. O...     »