Puluca, Nazan; Münsterer, Andrea; Prinzing, Anatol; Sexton, Zachary A; Lange, Rüdiger; Meyer-Saraei, Roza; Scharfschwerdt, Michael
Title:
In vitro comparison of everting vs. non-everting suture techniques for the implantation of a supra-annular biological heart valve.
Abstract:
Background: The aim of this study was to evaluate the hemodynamic effect of different suturing techniques for aortic valve replacement (AVR) in vitro. Whether or not the applied suturing technique impacts the outflow tract diameter by narrowing the annulus diameter was examined. Methods: The commonly applied non-everting pledget forced suture technique (NE, n=13) was compared with an everting pledget forced suture (ET, n=13) for AVR using the 25 mm St. Jude Trifecta aortic valve. Hemodynamic par...     »
Journal title abbreviation:
J Thorac Dis
Year:
2020
Journal volume:
12
Journal issue:
5
Pages contribution:
2443-2449
Fulltext / DOI:
doi:10.21037/jtd.2020.03.55
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/32642150