Benutzer: Gast  Login
Titel:

Improving metastable impact electron spectroscopy and ultraviolet photoelectron spectroscopy signals by means of a modified time-of-flight separation

Autor(en):
Spirkl, Florian M.; Kunz, Sebastian; Schweinberger, Florian F.; Farnbacher, Adrian N.; Schroeter, Richard; Heiz, Ulrich
Abstract:
The sepn. of UPS and metastable impact electron spectroscopy (MIES) is usually performed by a time-of-flight (ToF) sepn. using pre-set ToF for both types of signal. The authors present a new, improved ex situ signal sepn. method for the sepn. of MIES and UPS for every single measurement. Signal sepn. issues due to changes of system parameters can be overcome by changing the ToF sepn. and therefore allowing for the application of a wider range of measuring conditions. The method also enables to i...     »
Stichworte:
Metastable state Photoelectron spectrometers Time-of-flight mass spectrometers Time-of-flight mass spectrometry UV photoelectron spectra UV photoelectron spectroscopy (improving metastable impact electron spectroscopy and UPS signals by means of modified time-of-flight sepn.) Fluoropolymers Role: TEM (Technical or engineered material use), USES (Uses) (improving metastable impact electron spectroscopy and UPS signals by means of modified time-of-flight sepn. in app. contg.) Electron emission Ele...     »
Kongresstitel:
CAN 156:189423 Optical, Electron, and Mass Spectroscopy and Other Related Properties 1309-48-4 (Magnesium monoxide); 7439-98-7 (Molybdenum) Role: PEP (Physical, engineering or chemical process), PRP (Properties), PROC (Process) (improving metastable impact electron spectroscopy and UPS signals by means of modified time-of-flight sepn.); 7440-50-8 (Copper) Role: TEM (Technical or engineered material use), USES (Uses) (improving metastable impact electron spectroscopy and UPS signals by means of...     »
Zeitschriftentitel:
Rev. Sci. Instrum.
Jahr:
2012
Band / Volume:
83
Heft / Issue:
1
Seitenangaben Beitrag:
013114/1-013114/6
Volltext / DOI:
doi:10.1063/1.3677648
 BibTeX