Benutzer: Gast  Login
Titel:

Conduction Abnormalities and Pacemaker Implantations After SAPIEN 3 Vs SAPIEN XT Prosthesis Aortic Valve Implantation.

Dokumenttyp:
Comparative Study; Journal Article
Autor(en):
Husser, Oliver; Kessler, Thorsten; Burgdorf, Christof; Templin, Christian; Pellegrini, Costanza; Schneider, Simon; Kasel, Albert Markus; Kastrati, Adnan; Schunkert, Heribert; Hengstenberg, Christian
Abstract:
Transcatheter aortic valve implantation is increasingly used in patients with aortic stenosis. Post-procedural intraventricular conduction abnormalities and permanent pacemaker implantations remain a serious concern. Recently, the Edwards SAPIEN 3 prosthesis has replaced the SAPIEN XT. We sought to determine the incidences of new-onset intraventricular conduction abnormalities and permanent pacemaker implantations by comparing the 2 devices.We analyzed the last consecutive 103 patients undergoing transcatheter aortic valve implantation with SAPIEN XT before SAPIEN 3 was used in the next 105 patients. To analyze permanent pacemaker implantations and new-onset intraventricular conduction abnormalities, patients with these conditions at baseline were excluded. Electrocardiograms were recorded at baseline, after the procedure, and before discharge.SAPIEN 3 was associated with higher device success (100% vs 92%; P=.005) and less paravalvular leakage (0% vs 7%; P<.001). The incidence of permanent pacemaker implantations was 12.6% (23 of 183) with no difference between the 2 groups (SAPIEN 3: 12.5% [12 of 96] vs SAPIEN XT: 12.6% [11 of 87]; P=.99). SAPIEN 3 was associated with a higher rate of new-onset intraventricular conduction abnormalities (49% vs 27%; P=.007) due to a higher rate of fascicular blocks (17% vs 5%; P=.021). There was no statistically significant difference in transient (29% [20 of 69] vs persistent 19% [12 of 64]; P=.168) left bundle branch blocks (28% [19 of 69] vs 17% [11 of 64]; P=.154) when SAPIEN 3 was compared with SAPIEN XT.We found a trend toward a higher rate of new-onset intraventricular conduction abnormalities with SAPIEN 3 compared with SAPIEN XT, although this did not result in a higher permanent pacemaker implantation rate.
Zeitschriftentitel:
Rev Esp Cardiol (Engl Ed)
Jahr:
2016
Band / Volume:
69
Heft / Issue:
2
Seitenangaben Beitrag:
141-8
Sprache:
eng
Volltext / DOI:
doi:10.1016/j.rec.2015.06.024
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/26542175
TUM Einrichtung:
Klinik für Herz- und Kreislauferkrankungen im Erwachsenenalter (Prof. Schunkert)
 BibTeX