Biomechanical investigation of the primary stability of intramedullary compression nails in the proximal tibia: experimental study using interlocking screws in cryopreserved human tibias.

Document type:
Journal Article; Article
Author(s):
Gollwitzer, H; Karampour, K; Hauschild, M; Diehl, P; Busch, R; Mittelmeier, W
Abstract:
The biomechanical properties of the proximal tibia with respect to the application of compressible intramedullary nails were examined in a biomechanical model using human cadaver bones. A reference level was examined histomorphometrically in 24 cryopreserved human tibias. Three series of single or double interlocking fixations were tested at various distances from the tibial plateau. Yield strength, rigidity of the bone stock, maximum load, and compression distances were determined during axial...     »