Benutzer: Gast  Login
Titel:

Preventricular far-field sensing in the atrial channel of dual chamber pacemakers--an occasional cause of inappropriate mode switch.

Dokumenttyp:
Journal Article; Article
Autor(en):
Kolb, C; Aratma, S; Zrenner, B; Schmitt, C
Abstract:
INTRODUCTION AND AIMS OF THE STUDY: Atrial oversensing may trigger false positive mode switch to an asynchronous mode in dual chamber pacemakers. While myopotential oversensing and far-field R wave sensing within the postventricular atrial refractory period are well characterized, data about oversensing (near-field P wave or far-field R wave) within the atrioventricular delay is limited. Aim of the study was to determine the incidence of preventricular oversensing in the atrial channel of curren...     »
Zeitschriftentitel:
J Interv Card Electrophysiol
Jahr:
2004
Band / Volume:
10
Heft / Issue:
3
Seitenangaben Beitrag:
231-5
Sprache:
eng
Volltext / DOI:
doi:10.1023/B:JICE.0000026917.82516.eb
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/15133360
Print-ISSN:
1383-875X
TUM Einrichtung:
I. Medizinische Klinik und Poliklinik (Kardiologie)
 BibTeX