Combining perimodiolar electrode placement and atraumatic insertion properties in cochlear implantation -- fact or fantasy?
Abstract:
CONCLUSIONS: Except for basal cochlear traumatization, all specimens implanted into scala tympani showed atraumatic insertion properties and good perimodiolar electrode positioning. Cochleostomy preparation and placement can have a significant impact on levels of basal cochlear trauma. OBJECTIVE: In the past, perimodiolar cochlear implant electrodes increased the risk for intracochlear traumatization when compared to free-fitting arrays. Recently, however, clinical evidence for atraumatic perimo...     »
Journal title abbreviation:
Acta Otolaryngol
Year:
2006
Journal volume:
126
Journal issue:
5
Pages contribution:
475-82