User: Guest  Login
Document type:
Journal Article
Author(s):
Heiser, Clemens
Title:
Advanced titration to treat a floppy epiglottis in selective upper airway stimulation.
Abstract:
The following report presents a case of a patient with high residual apnea-hypopnea index at 6 months following implantation of an upper airway stimulation device who improved significantly after advanced titration via drug-induced sedated endoscopy by changing the electrode configuration for stimulation from bipolar to monopolar. This case demonstrates the utility of postsurgical advanced titration to optimize programming settings for improved clinical outcomes. Laryngoscope, 126:S22-S24, 2016.
Journal title abbreviation:
Laryngoscope
Year:
2016
Journal volume:
126 Suppl
Pages contribution:
S22-4
Language:
eng
Fulltext / DOI:
doi:10.1002/lary.26118
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/27480503
Print-ISSN:
0023-852X
TUM Institution:
Hals-Nasen-Ohrenklinik und Poliklinik
 BibTeX