Benutzer: Gast  Login
Mehr Felder
Einfache Suche
Titel:

Evaluation of the SUV values calculation and 4D PET integration in the radiotherapy treatment planning system.

Dokumenttyp:
Journal Article
Autor(en):
Jacob, V; Astner, ST; Bundschuh, RA; Busch, R; Souvatzoglou, M; Wendl, C; Kneschaurek, P; Grosu, AL
Abstract:
To evaluate the SUV calculation and integration of the gated (4D) PET in the iPlan 4.0 treatment planning software (BrainLAB).Phantom and patient data for different tracers were used. Two comparisons were performed for each patient: for the delineated VOI, the maximum value of SUV in iPlan was compared with the results from TrueD software. For 10 patients lesion volumes were defined in both systems for a given SUV threshold and differences were calculated. For four patients examined with respira...     »
Zeitschriftentitel:
Radiother Oncol
Jahr:
2011
Band / Volume:
98
Heft / Issue:
3
Seitenangaben Beitrag:
323-9
Sprache:
eng
Volltext / DOI:
doi:10.1016/j.radonc.2011.01.001
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/21296443
Print-ISSN:
0167-8140
TUM Einrichtung:
Institut für Medizinische Statistik und Epidemiologie; Klinik und Poliklinik für Nuklearmedizin; Klinik und Poliklinik für RadioOnkologie und Strahlentherapie
 BibTeX