Benutzer: Gast  Login

Titel:

The dynamic locking screw (DLS) can increase interfragmentary motion on the near cortex of locked plating constructs by reducing the axial stiffness.

Dokumenttyp:
Journal Article; Article
Autor(en):
Döbele, S; Horn, C; Eichhorn, S; Buchholtz, A; Lenich, A; Burgkart, R; Nüssler, AK; Lucke, M; Andermatt, D; Koch, R; Stöckle, U
Abstract:
BACKGROUND: The plate-screw interface of an angular stable plate osteosynthesis is very rigid. So far, all attempts to decrease the stiffness of locked plating construct, e.g. the bridged plate technique, decrease primarily the bending stiffness. Thus, the interfragmentary motion increases only on the far cortical side by bending the plate. To solve this problem, the dynamic locking screw (DLS) was developed. MATERIALS AND METHODS: Comparison tests were performed with locking screws (LS) and DLS...     »
Zeitschriftentitel:
Langenbecks Arch Surg
Jahr:
2010
Band / Volume:
395
Heft / Issue:
4
Seitenangaben Beitrag:
421-8
Sprache:
eng
Volltext / DOI:
doi:10.1007/s00423-010-0636-z
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/20358382
Print-ISSN:
1435-2443
TUM Einrichtung:
Klinik und Poliklinik für Orthopädie und Sportorthopädie; Klinik und Poliklinik für Unfallchirurgie