User: Guest  Login
More Searchfields
Simple search
Title:

Influence of test temperature on biomechanical properties of all-inside meniscal repair devices and inside-out meniscus sutures--evaluation of an isolated distraction loading, worst-case scenario.

Document type:
Journal Article; Research Support, Non-U.S. Gov't
Author(s):
Brucker, PU; Favre, P; Puskas, GJ; von Campe, A; Koch, PP; Meyer, DC
Abstract:
Studies on the biomechanical properties of meniscus repairs are usually performed at room instead of body temperature. However, various all-inside meniscal repair devices include bioabsorbable materials, which are mechanically sensitive to higher environmental temperatures. Therefore, we hypothesize that current test standards may systematically lead to a false overestimation of their performance.In 84 cadaveric bovine lateral menisci, an artificial vertical lesion was repaired with different al...     »
Journal title abbreviation:
Clin Biomech (Bristol, Avon)
Year:
2011
Journal volume:
26
Journal issue:
7
Pages contribution:
749-53
Language:
eng
Fulltext / DOI:
doi:10.1016/j.clinbiomech.2011.02.007
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/21570754
Print-ISSN:
0268-0033
TUM Institution:
Fachgebiet Sportorthopädie (Prof. Imhoff)
 BibTeX