User: Guest  Login
More Searchfields
Simple search
Title:

Multicenter evaluation of a new automated electrochemiluminescence immunoassay for the quantification of testosterone compared to liquid chromatography tandem mass spectrometry.

Document type:
Journal Article
Author(s):
Brandhorst, G; Streit, F; Kratzsch, J; Schiettecatte, J; Roth, HJ; Luppa, PB; Körner, A; Kiess, W; Binder, L; Oellerich, M; von Ahsen, N
Abstract:
A new electrochemiluminescence testosterone immunoassay (Elecsys® Testo II) was evaluated.Within-run precision, total precision, and method comparisons to the former immunoassay and LC-MS/MS were investigated. Reference intervals were established for adults and children.Imprecision ranged from 0.9% to 8.5%. Method comparisons revealed slopes from 0.68 to 1.32 and coefficients of correlation from 0.893 to 0.991.Improved agreement with LC-MS/MS was demonstrated, especially for female patients. The...     »
Journal title abbreviation:
Clin Biochem
Year:
2011
Journal volume:
44
Journal issue:
2-3
Pages contribution:
264-7
Language:
eng
Fulltext / DOI:
doi:10.1016/j.clinbiochem.2010.09.024
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/20923677
Print-ISSN:
0009-9120
TUM Institution:
Institut für Klinische Chemie und Pathobiochemie
 BibTeX