User: Guest  Login
More Searchfields
Simple search
Document type:
Journal Article; Research Support, Non-U.S. Gov't; Article
Author(s):
Jilek, C; Tzeis, S; Vrazic, H; Semmler, V; Andrikopoulos, G; Reents, T; Fichtner, S; Ammar, S; Rassias, I; Theodorakis, G; Weber, S; Hessling, G; Deisenhofer, I; Kolb, C
Title:
Safety of screening procedures with hand-held metal detectors among patients with implanted cardiac rhythm devices: a cross-sectional analysis.
Abstract:
Case reports suggest that the hand-held metal detectors used for security screening generate electromagnetic fields that may interfere with pacemaker and implantable cardioverter-defibrillator (ICD) function.To assess changes in function of pacemakers and ICDs after exposure to hand-held metal detectors.Cross-sectional study.Two medical centers in Europe.388 patients (209 with pacemakers and 179 with ICDs) presenting for routine follow-up of device function between September 2009 and December 20...     »
Journal title abbreviation:
Ann Intern Med
Year:
2011
Journal volume:
155
Journal issue:
9
Pages contribution:
587-92
Language:
eng
Fulltext / DOI:
doi:10.1059/0003-4819-155-9-201111010-00005
Pubmed ID:
http://view.ncbi.nlm.nih.gov/pubmed/22041947
Print-ISSN:
0003-4819
TUM Institution:
I. Medizinische Klinik und Poliklinik (Kardiologie)
 BibTeX