Titel:
Safety of screening procedures with hand-held metal detectors among patients with implanted cardiac rhythm devices: a cross-sectional analysis.
Abstract:
Case reports suggest that the hand-held metal detectors used for security screening generate electromagnetic fields that may interfere with pacemaker and implantable cardioverter-defibrillator (ICD) function.To assess changes in function of pacemakers and ICDs after exposure to hand-held metal detectors.Cross-sectional study.Two medical centers in Europe.388 patients (209 with pacemakers and 179 with ICDs) presenting for routine follow-up of device function between September 2009 and December 20...     »