Benutzer: Gast  Login
Titel:

Point-of-care determination of neonatal bilirubin with the blood gas analyzer RapidLab 1265.

Dokumenttyp:
Journal Article; Research Support, Non-U.S. Gov't
Autor(en):
Mielsch, C; Zimmermann, A; Wagner, D; Matthes, B; Schlebusch, H; Luppa, PB
Abstract:
The aim of the study was to evaluate the comparability of the new neonatal bilirubin method on the RapidLab 1265 blood gas analyzer. This point-of-care testing (POCT) device has the option for the determination of neonatal bilirubin, making it potentially valuable for use in neonate intensive care units or in outpatient ambulances.We paired 240 patient samples for intermethod comparisons between the new POCT method and the routine laboratory method (Vitros 350 chemistry system with BuBc slide)....     »
Zeitschriftentitel:
Clin Chem Lab Med
Jahr:
2010
Band / Volume:
48
Heft / Issue:
10
Seitenangaben Beitrag:
1455-61
Sprache:
eng
Volltext / DOI:
doi:10.1515/CCLM.2010.279
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/20604721
Print-ISSN:
1434-6621
TUM Einrichtung:
Institut für Klinische Chemie und Pathobiochemie