Original title:
Improving Quantitative Imaging Accuracy for Dosimetry of Lutetium-177-PSMA Imaging &Therapy
Translated title:
Verbesserung der quantitativen Bildgebungsgenauigkeit für die Dosimetrie von Lutetium-177-PSMA Imaging & Therapy
Author:
Karimi Ghodoosi, Elham
Year:
2021
Document type:
Dissertation
Faculty/School:
Fakultät für Medizin
Advisor:
Ziegler, Sibylle (Prof. Dr.)
Referee:
Ziegler, Sibylle (Prof. Dr.); Navab, Nassir (Prof. Dr.)
Language:
en
Subject group:
MED Medizin
Keywords:
(177)Lu; Attenuation map; Calibration factor; SPECT/CT; Scatter fraction; Scatter window
TUM classification:
MED 540
Abstract:
The objective of this study was to evaluate the image degrading factors in quantitative 177Lu SPECT imaging. Overall, we found that for an accurate quantitative 177Lu SPECT imaging, the size of the calibration phantom should be considered at the calibration process. There is a considerable influence of the attenuation map, scatter correction method, and scatter window width on the SPECT/CT calibration reproducibility. Due to overlapping, special care needs to be taken in organ delineation if pla...     »
Translated abstract:
Das Ziel dieser Studie war es, die bildverschlechterenden Faktoren bei der quantitativen 177Lu SPECT-Bildgebung zu bewerten. Insgesamt haben wir festgestellt, dass für eine genaue quantitative 177Lu SPECT-Bildgebung die Größe des Kalibrierungsphantoms beim Kalibrierungsprozess berücksichtigt werden sollte. Es gibt einen erheblichen Einfluss der Dämpfungskarte, der Streukorrekturmethode und der Streufensterbreite auf die Reproduzierbarkeit der SPECT / CT-Kalibrierung. Aufgrund von Überlappungen i...     »
WWW:
https://mediatum.ub.tum.de/?id=1574486
Date of submission:
30.12.2020
Oral examination:
28.07.2021
File size:
2809349 bytes
Pages:
107