Chen, X.; Wang, L.; Fallavollita, P.; Navab, N.
Titel:
Error analysis of the x-ray projection geometry of camera-augmented mobile C-arm
Abstract:
The Camera-Augmented Mobile C-arm (CamC) augments X-ray by camera images and is calibrated in a suboptimal way for X-ray to be warped onto optical images in order to align them and have common projection geometry. We investigated and compared calibration and distortion correction accuracy of: (i) the existing CAMC method (ii) Zhang's method for optical cameras, and (iii) traditional C-arm method. The RMS errors based on a synthetic model and pose estimation shows that the traditional C-arm metho...     »