Benutzer: Gast  Login
Dokumenttyp:
English Abstract; Journal Article
Autor(en):
Israel, CW; Bänsch, D; Böcker, D; Butter, C; Chun, J; Deisenhofer, I; Eckardt, L; Geller, JC; Hanke, T; Klingenheben, T; Piorkowski, C; Schumacher, B
Titel:
[Recommendations of the Working Group of Arrhythmias of the German Society of Cardiology on the approach to patients with Riata® and Riata ST® leads (St. Jude Medical). Nucleus of the Working Group of Arrhythmias of the German Society of Cardiology].
Abstract:
Riata® and Riata ST® implantable cardioverter defibrillator (ICD) leads (St. Jude Medical, Sylmar, CA) show an increased incidence of insulation defects, particularly "inside-out" lead fracture where inner, separately insulated cables penetrate through the surrounding silicone of the lead body. The exact incidence of Riata® lead problems is not clear and seems to range between 2-4% per year in the first 5 years after implantation according to new registry data. We recommend beyond a detailed inf...     »
Zeitschriftentitel:
Herzschrittmacherther Elektrophysiol
Jahr:
2012
Band / Volume:
23
Heft / Issue:
2
Seitenangaben Beitrag:
107-15
Sprache:
de
Volltext / DOI:
doi:10.1007/s00399-012-0186-x
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/22847674
Print-ISSN:
0938-7412
TUM Einrichtung:
Klinik für Herz- und Kreislauferkrankungen im Erwachsenenalter (Prof. Schunkert)
 BibTeX