Abstract:
Riata® and Riata ST® implantable cardioverter defibrillator (ICD) leads (St. Jude Medical, Sylmar, CA) show an increased incidence of insulation defects, particularly "inside-out" lead fracture where inner, separately insulated cables penetrate through the surrounding silicone of the lead body. The exact incidence of Riata® lead problems is not clear and seems to range between 2-4% per year in the first 5 years after implantation according to new registry data. We recommend beyond a detailed inf...    »
 