Benutzer: Gast  Login
Dokumenttyp:
Evaluation Studies; Journal Article; Multicenter Study; Article
Autor(en):
Alber, B; Hein, R; Garbe, C; Caroli, U; Luppa, PB
Titel:
Multicenter evaluation of the analytical and clinical performance of the Elecsys S100 immunoassay in patients with malignant melanoma.
Abstract:
The aim of this multicenter study was to evaluate the technical and clinical performance of the Elecsys S100 electrochemiluminescence immunoassay and to assess its utility as a tumor marker in patients with malignant melanoma. Imprecision studies yielded within-run coefficients of variation (CVs) of 0.7-2.0% and between-day CVs of 1.0-6.4%. Serum samples that were distributed to 11 participating laboratories for a comparability analysis resulted in excellent recoveries of 93-105% related to the...     »
Zeitschriftentitel:
Clin Chem Lab Med
Jahr:
2005
Band / Volume:
43
Heft / Issue:
5
Seitenangaben Beitrag:
557-63
Sprache:
eng
Volltext / DOI:
doi:10.1515/CCLM.2005.097
PubMed:
http://view.ncbi.nlm.nih.gov/pubmed/15899680
Print-ISSN:
1434-6621
TUM Einrichtung:
Institut für Klinische Chemie und Pathobiochemie; Klinik und Poliklinik für Dermatologie und Allergologie
 BibTeX